- Complies with ATEX II 1G Ex ia IIC T4 requirements.
- Built-in pressure / vacuum hand pump With fine vernier And bleeder valve
- Ranges of 30, 100 and 300 psi (2, 7 and 20 bar).
- Pressure measurement to 0.05% of full span. Using an internal pressure sensor
- Pressure measurement to 200 bar using one of eight intrinsically safe Fluke 700PEx pressure modules.
- Multiple selectable pressure units
- Current measurement with 0.02% accuracy and 0.001 mA resolution.
- Pressure switch test function
- Min-max display function and hold
- Compact and lightweight
- Intuitive, push-button user interface
- Durable and reliable Suitable for field use
- Measurement accuracy
- Pressure
Fluke 718Ex 30G
Range: -830 mbar to 2 bar (internal sensor)
Overpressure: 3x FS
Resolution: 0.0001 bar
Accuracy: 0.05% FS
Function: zero, minimum, maximum, hold display, damper
Note: For use with non-corrosive gases.
Fluke 718Ex 100G
Range: -830 mbar to 7 bar (internal sensor)
Overpressure: 2x FS
Resolution: 0.001 bar
Accuracy: 0.05% FS
Function: zero, minimum, maximum, hold display, damper
Note: For use with non-corrosive gases.
Fluke 718Ex 300G
Range: -830 mbar to 20 bar (internal sensor)
Pressure is too high: 375 psi, 25 bar.
Resolution: 0.0001 bar
Accuracy: 0.05% FS